TRANSESOPHAGEAL CARDIAC STIMULATOR MODEL 7A WITH PREAMPLIFIER MODEL 3 FOR ECG RECORDING
CardioCommand Model 3 Recording Preamplifier
Electrical Amplifier: Input impedance greater than 500k ohms at 10 Hz, Differential input gain of 2
Physical
|
Size: |
2.375`` (6 cm) high x 6.375`` (16.2 cm) deep x 5.125`` (13 cm) wide |
|
Weight: |
1 lb (0.45 kg) without batteries |
|
Material: |
Impact-resistant ABS plastic enclosure |
Power Supply
|
Battery: |
9-volt (2). The unit should operate normally for 3 years with fully charged batteries at the specified operating conditions. |
|
Warning: |
Use of batteries other than those specified by the manufacturer may result in the device not operating within the stated specifications. |
CardioCommand Model 7A Transesophageal Cardiac Stimulator Electrical Pulse AMPLITUDE: Continuously variable, 0-40 mA Pulse DURATION: Continuously variable rectangular pulse, 0-10 ms Pacing RATE: Continuously variable 60-300 BPM (X1), 120-600 BPM (X2) Load impedance: Constant current at impedances up to 4k ohms at 20 mA. Rate Limit: 343 BPM (runaway protection). Mode of operation in the presence of sensed interference: Shutdown. Accuracy: ± 10 % of full scale.
Physical
|
Size: |
2.63`` (6.7 cm) high x 7.75`` (19.7 cm) deep x 8.13`` (20.6 cm) wide |
|
Weight: |
2.25 lbs (1 kg) |
|
Material: |
Impact-resistant ABS plastic enclosure | Power Supply
|
Battery type: |
9-volt (4). The unit should operate normally for 24 hours with fully charged batteries at the specified operating conditions. The device should operate normally for 2 hours upon activation of the ``low battery`` indicator. |
|
Warning: |
Use of batteries other than those specified by the manufacturer may result in the device not operating within the stated specifications.
|
BIPOLAR ESOPHAGEAL PACING/RECORDING CATHETERS
CardioCommand Product Specifications Pacing/Recording Esophageal Catheters
TAPSCOPE 210TM
|
Electrode configuration: |
Bipolar, stainless steel |
|
Wiring: |
26 AWG copper |
|
Size: |
10 French, length 96 cm |
|
Electrode length: |
10 mm |
|
Electrode spacing: |
20 mm center-to-center |
|
Stylet: |
Stainless steel (optional, 210-S) |
|
Catheter material: |
USP Medical Grade Class VI PVC |
|
Catheter printing: |
Depth of insertion from teeth (cm) |
|
Packaging: |
1 device/pouch, 10 pouches/carton |
TAPCATH 205TM
|
Electrode configuration: |
Bipolar, stainless steel |
|
Wiring: |
Copper |
|
Size: |
5 French, length 100 cm |
|
Electrode length: |
3.5 mm |
|
Electrode spacing: |
13 mm center-to-center |
|
Stylet: |
Stainless steel |
|
Catheter material: |
USP Medical Grade PVC |
|
Catheter markings: |
Depth of insertion from teeth (cm) |
|
Packaging: |
1 device/pouch, 5 pouches/carton |
|


|